IVDR transitional provisions

This webpage provides information for industry stakeholders about the In Vitro Diagnostic Medical Devices Regulation (IVDR) transitional provisions.

What are transitional provisions?

To support the supply of IVDs to the EU market, Articles 110 and 113 of the IVDR set out a number of transitional provisions (TPs).

TPs give more time to manufacturers of certain devices already placed on the market to comply with the IVDR.

When do transitional provisions apply?

On 26 May 2022, the IVDR replaced the in vitro diagnostics directive (IVDD).

TPs may apply to IVDs CE marked in line with the IVDD where there are no significant changes in the design and intended purpose.

IVDs ‘up classified’ under the IVDR can also avail of TPs once the manufacturer has drawn up a declaration of conformity before 26 May 2022.

Although some IVDs may be subject to TPs, IVDR requirements relating to post-market surveillance, market surveillance, vigilance, registration of economic operators and of devices, are fully applicable from 26 May 2022.

Article 112 of the IVDR calls out exceptions to some key TPs relating to Eudamed.

Change to IVDR transitional provisions

In January 2022, the European Commission adopted Regulation 2022/112. Amongst other things, this changed the transitional timelines for devices set out under Article 110 of the IVDR.

We have provided a summary of the key information outlined in this amendment below.

IVDR classificationMay be placed on the market until *Can continue to be made available until *
Class D26 May 202526 May 2026
Class C26 May 202626 May 2027
Class B26 May 202726 May 2028
Class A (sterile)26 May 202726 May 2028

IVDs with an IVDD notified body certificate

These certificates will continue to be valid up until 26 May 2025 at the latest. Economic operators may continue to make available or put into service devices with a certificate issued before 26 May 2022 until 26 May 2025.

IVDs without a notified body certificate issued under the IVDD

Class A non-sterile IVDs must be IVDR compliant by 26 May 2022.

Some IVDs did not require notified body oversight under the IVDD but will require it under the IVDR. These IVDs can remain on the market after 26 May 2022 only if the manufacturer drew up the IVDD declaration of conformity before this date.

The transition times for these IVDs depend on their IVDR classification as described in table 1 above.

When do I need to place a UDI carrier on my IVD?

Article 24(4) notes that manufacturers must place UDI carriers on the label of IVDs and on all higher levels of packaging.

Article 113 of the IVDR sets out a number of provisions and covers the staggered application of device labelling and UDI carriers. We have provided a summary of UDI timelines in the table below.

IVDR classificationDeadline to include UDI on the device label
Class D26 May 2023

Class C

Class B

26 May 2025
Class A26 May 2027

Transitional provisions for in-house manufactured devices

Regulation 2022/112 introduces a phased application of the requirements for IVDs made and used within health institutions.

We provide a summary of these timelines in the table below. You should read these timelines together with Article 5(5) of the IVDR.

IVDR referenceDescriptionApplicable from
Article 5(5) opening paragraph & Annex IAnnex I applies to in-house manufactured devices.26 May 2022
Article 5(5)b,c & e-iSee Article 5(5) for more details.26 May 2024
Article 5(5)dRequirement to provide justification that a patient group's specific needs cannot be met, or cannot be met at the appropriate level of performance by an equivalent device available on the market. 26 May 2028
Article 5(5) final paragraphMember States' rights to request information/restrict manufacture and use of devices and inspect health institutions.26 May 2022

Transitional provisions for class D IVDs

The Medical Devices Coordination (MDCG) has developed guidance on the application of transitional provisions for certification of class D IVDs under the IVDR. This document provides information for IVD manufacturers on how to apply IVDR provisions related to expert panels and European Union reference laboratories (EURLs).